beautypg.com

Alarm time settings – AASTRA 600c_d for SIP-DECT User Manual EN User Manual

Page 42

background image

36

• No movement alarm
• Escape alarm

All three alarm types can be activated in parallel. The alarm number is the same for all alarm types. If
no alarm number is stored, the SOS emergency number of the emergency call key / SOS key is used
instead.
The alarm to be signalled is always the first alarm type to be recognised. Activated sensor alarms are
displayed using the

symbol on the first line of the display. To check which alarm is activated, use

the

Active features menu (among others).

Make sure the handset is worn as freely as possible (on a belt clip, carry-strap, etc.), i.e. not in a
pocket, to ensure that the feature is able to operate reliably.

Mandown
The handset is normally worn more or less vertically, e.g. attached directly to the wearer’s belt clip. A
position check is carried out continuously to verify that the handset is still in the vertical position. If
the handset is detected as being at a certain (adjustable) angle from the vertical, an alarm is trigge-
red after a predefinable amount of time as this may indicate an emergency.

No movement alarm
This alarm is triggered if the handset has been completely still for a user-definable amount of time
(no movement by the handset or the person carrying the handset). The sensor has three sensitivity
settings. To guarantee a reliable operation it is advisable to test the sensitivity of the no movement
alarm with the person to be monitored. The no movement alarm is switched off whenever the hand-
set is being charged in the charger cradle or via the USB socket.

Escape alarm
The escape alarm is triggered if the sensor detects violent handset movements for user-definable
period of time. As with the no movement alarm the sensor has three sensitivity settings. To guaran-
tee a reliable operation it is advisable to test the sensitivity of the escape alarm with the person to
be monitored.
There is a setting for each alarm type to activate / deactivate the sensor supervision in an active call
(Alarm in conversation).

Alarm time settings

All the settings can be found in the menu ❯❯❯

Protection > Alarm Sensor

The alarm times are structured as follows.

Delay
This is the period of time that begins after the triggering event (“mandown”, no movement, escape).
If during this time the triggering event is withdrawn for at least 2 seconds, the delay begins from the

Basic functions