Compensation entry methods, Softkey brings up the, Form. f – ACU-RITE VRO 300M User Manual
Page 56: You can change only the, Field. use the, Softkey. the

VRO 300M
Setup
49
Compensation Entry Methods
You can enter the segment boundaries and compensation factors using the
keypad, or you can have the readout calculate them for you. Normally, you
would use the keypad entry method only to restore or adjust a known
compensation value.
Manual Entry
This brings up the softkeys needed to let you enter compensation
and segment information using the keypad.
The
NEW
SEGMENT
softkey brings up the
DEFINE SEGMENT
form.
F
ACTOR
—Enter the compensation
factor in parts per million.
S
EGMENT
—You can change only the
TO
field. Use the
ENCODER
END
softkey if the
boundary is at the end of the
encoder. Use the
POINT
softkey to
enter the physical location of the
segment boundary. The current
reading head location may be entered using
TEACH
POSITION
softkey.
The
DONE
softkey saves the segment information in the table.
Continue adding segments with the
NEW
SEGMENT
softkey. You must add them in
order, from the encoder’s most negative end to its most positive end.
The
CLEAR
TABLE
softkey will delete the information from the table, erasing all
segments and compensation factors, so that the readout will not perform any
error compensation.
ENTER
COMP
FACTOR
PPM
2
DEFINE SEGMENT
SEGMENT
FROM
TO
ENCODER END
ENCODER END
NEW
SEGMENT
CLEAR TABLE
ANCHOR
SEGMENT 1
CANCEL
DONE