background image

20

SZ2

series

Fig. 18

Fig. 17

Push up.

Clamp.

To increase

To decrease

Clamping screw mount holes

Stage

Load capacity: 1 kg

Clamping knob

Focus adjustment knobs

Stroke: Approx. 21 mm
Rotation tension adjustable.

1. Using the provided screw and Allen wrench, attach the SZ2-FO up/down moving stage into the screw hole on the stage

plate mount hole of an applicable base or that on the stage plate (SZ-STL).
The SZ2-FO can also be installed so that the focus adjustment knobs are located on the near side of the operator. This
orientation is obligatory when the IHE holder for the LSGA oblique illumination system is used.

2. Loosen the clamping knob on the stage, attach the stage plate, SZH-SG or SZH-SC, and tighten the clamping knob

again.

Removing the stage plate (Fig. 17)

To remove the stage plate, loosen the clamping knob @ and push up
the stage plate from below it.

2

External View and Nomenclature

3

Installation

4

Operation

@

Adjusting the focus adjustment knob tension (Fig. 18)

}This operation is intended to facilitate the rotation of the knobs while

preventing spontaneous drop of the up/down moving stage. It is recom-
mended to set the knob tension to a slightly higher level than the point
where spontaneous drop occurs.
a. Hold the left and right focus adjustment knobs @ with both hands, fix

the left knob and rotate the right knob. The rotation tension of the
knobs increases or decreases according to the direction in which the
right knob is rotated.

b. If the rotation tension is increased too much, accurate focusing is not

possible and the mechanism may be damaged.

@