beautypg.com

Warning, High altitude dives and personal adjustment, Altitude adjustment – SUUNTO Gekko User Manual

Page 36

background image

34

- the OLF bar graph reaches 100%.

The blinking of the segments exceeding 80% will stop, when the OLF

is not loading anymore. At that point the PO

2

is less than 0.5 bar.

Continuous beeps for 3 minutes occur, when:
- the set oxygen partial pressure limit is exceeded. The maximum depth is

replaced with a current blinking PO

2

value. You should immediately ascend

above the PO

2

depth limit (Fig. 3.23.).

WARNING!

WHEN THE OXYGEN LIMIT FRACTION INDICATES THAT THE MAXI-

MUM LIMIT IS REACHED, YOU MUST IMMEDIATELY ASCEND UNTIL

THE WARNING STOPS BLINKING! Failure to take action to reduce oxygen

exposure after the w arning is gi ven can rapidly increase the risk of oxygen

toxicity and the risk of injury or death.

3.7. HIGH ALTITUDE DIVES AND PERSONAL
ADJUSTMENT

The dive computer can be adjusted both for diving at altitude and also to increase

the conservatism of the mathematical nitrogen model.

3.7.1. Altitude Adjustment

When programming the instrument for the correct altitude, you need to select the

correct Altitude Mode according to Table 3.4. The dive computer will adjust its

mathematical model according to the entered altitude mode, gi ving shorter no-

decompression times at higher altitudes (see Section 6.1. “Operating Principles”,

Table 6.1. and 6.2.).
The entered Altitude Adjustment Mode is indicated by mountain symbols (A0, A1

TABLE 3.4. ALTITUDE ADJUSTMENT RANGES

Altitude

mode

Symbol

on display

Altitude

range

A0

A1

A2

0 - 300 m [0 - 1000 ft]

300 - 1500 m [1000 - 5000 ft]

1500 - 3000 m [5000 - 10000]

= one mountain, or A2 = two mountains). Section 4.2.4. “ Personal Adjustments”

describes how the Altitude Mode is adjusted.
Traveling to a higher elevation can temporarily cause a change in the equilibrium