1740A/1750A Series Waveform/Vector Monitor User Manual

A-1

Appendix A

Performance Specification

The specification tables use the following abbreviations:

Performance Requirement

(Req)

.

All

performance requirements in the specification tables are
identified with

Req, and can be assured by completing

the Performance Check Procedure (1740A/
1750A/1760--Series Service Manual.) Allow a warm-up
time of 20 minutes.

Performance requirements in the electrical specifications
are valid over an ambient temperature range of +20

° C to

+30

° C. The Performance Requirement tolerances listed

in the Electrical Specification are doubled over the
temperature range of 0

° C to +50° C, unless otherwise

specified. Test equipment used to verify Performance
Requirements must be calibrated and working within the
limits specified in the Service Manual Equipment Re-
quired List.

Reference Information.

Information that ampli-

fies a performance requirement or is of special impor-
tance is indicated by

RI. There is no need to check these

items to a specific tolerance.

Reliability

Mean Time Between Failure (MTBF)

The demonstrated MTBF goal is 40637 hours MTBF at
25

° C ambient, which will be monitored from field

failure reports.

Mean Time to Repair (MTTR)