Technical Specifications

En

glish

Guidance and manufacturer’s declaration – electromagnetic immunity (Continued)

The VPAP Malibu is intended for use in the electromagnetic environment specified below. The customer or

the user of the VPAP Malibu should assure that it is used in such an environment.

Immunity
test

IEC60601-1-2
test level

Compliance
level

Electromagnetic environment –
guidance

Portable and mobile RF communications

equipment should be used no closer to any

part of the VPAP Malibu, including cables,

than the recommended separation distance

calculated from the equation applicable to the

frequency of the transmitter.

Recommended separation distance

Conducted RF

IEC 61000-4-6

3 Vrms

150 kHz to 80

MHz

3 Vrms

d = 1.17

P

Radiated RF

IEC 61000-4-3

10 V/m

80 MHz to 2.5

GHz

10 V/m

d = 0.35

P 80 MHz to 800 MHz

d = 0.70

P 800 MHz to 2.5 GHz

where P is the maximum output power rating

of the transmitter in watts (W) according to

the transmitter manufacturer and d is the

recommended separation distance in meters

(m).

Field strengths from fixed RF transmitters, as

determined by an electromagnetic site
survey,

a

should be less than the compliance

level in each frequency range.

b

Interference may occur in the vicinity of

equipment marked with the following symbol:

NOTE 1: At 80 MHz and 800MHz, the higher frequency range applies.

NOTE 2: These guidelines may not apply in all situations. Electromagnetic propagation is affected by

absorption and reflection from structures, objects and people.

a

Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless) telephones and

land mobile radios, amateur radio, AM and FM radio broadcast and TV broadcast cannot be predicted

theoretically with accuracy. To assess the electromagnetic environment due to fixed RF transmitters, an

electromagnetic site survey should be considered. If the measured field strength in the location in which

the VPAP Malibu is used exceeds the applicable RF compliance level above, the VPAP Malibu should be

observed to verify normal operation. If abnormal performance is observed, additional measures may be

necessary, such as reorienting or relocating the VPAP Malibu.

b

Over the frequency range 150 kHz to 80 MHz, field strengths should be less than 10 V/m.