Technical Specifications

En

glish

Technical Specifications

Performance
Operating pressure range: 3 to 25 cm H

2

O

Maximum single fault pressure: 40 cm H

2

O

Pressure measurement tolerance: ±0.5 cm H

2

O + 4% of the measured reading

Flow measurement tolerance: ±0.2 L/s
Dynamic pressure characteristics
VSet mode:
4 to 25 cm H

2

O (measured at the end of standard 6’6" air tubing);

AutoSet Max pressure: 20 cm H

2

O; Pressure Support: 0 to 10 cm H

2

O

S mode: IPAP: 3 to 25 cm H

2

O (measured at the end of standard 6’6" air tubing);

EPAP: 3 to 25 cm H

2

O (measured at the end of standard 6’6" air tubing)

CPAP mode: 4 to 20 cm H

2

O (measured at the end of standard 6’6" air tubing)

Dynamic flow characteristics
130 L/min at 3 to 25 cm H

2

O

Sound pressure level: <30 dB (tested in accordance with the requirements of
ISO 17510-1:2002)
Data storage: 365 days of usage, leak, pressure, AHI and AI data
Dimensions (L x W x H): 10.6" x 9.1" x 5.6"
Weight: 5.1lb
Power supply: Input range 100–240V, 50–60Hz, 40VA (typical power
consumption), < 100VA (maximum power consumption)
Housing construction: Flame retardant engineering thermoplastic
Environmental conditions
Operating temperature: 41

o

F to 104

o

F (+5

o

C to +40

o

C)

Operating humidity: 10–95% non-condensing
Storage and transport temperature: -4

o

F to 140

o

F (-20

o

C to +60

o

C)

Storage and transport humidity: 10–95% non-condensing
Electromagnetic Compatibility
Product complies with all applicable electromagnetic compatibility requirements
(EMC) according to IEC60601-1-2, for residential, commercial, and light industry
environments. For further details, see “Guidance and Manufacturer’s
Declaration - Electromagnetic Emissions and Immunity” on page 37.
Air filter: Two-layered, powder-bonded, polyester non-woven fiber
Air tubing: Flexible plastic, 1 x 6’6" (2 m) or 1 x 9’10" (3 m) length
Air outlet: 22mm taper, compatible with EN 1281-1:1997 Anaesthetic &
Respiratory Equipment - Conical Connectors
IEC 60601-1 classification: Class II (double insulated), Type CF
Note: The manufacturer reserves the right to change these specifications
without notice.

268106r1.book Page 35 Thursday, November 30, 2006 2:35 PM