37
Technical Specifications
En
glish
Guidance and Manufacturer’s Declaration - Electromagnetic
Emissions and Immunity
Guidance and manufacturer’s declaration – electromagnetic emissions
The VPAP Malibu is intended for use in the electromagnetic environment specified below. The
customer or the user of the VPAP Malibu should assure that it is used in such an environment.
Emissions test
Compliance
Electromagnetic environment -
guidance
RF emissions CISPR 11
Group 1
The VPAP Malibu uses RF energy only
for its internal function. Therefore, its
RF emissions are very low and are not
likely to cause any interference in
nearby electronic equipment.
RF emissions CISPR 11
Class B
The VPAP Malibu is suitable for use in
all establishments, including domestic
establishments and those directly
connected to the public low-voltage
network that supplies buildings used
for domestic purposes.
Harmonic Emissions
IEC 61000-3-2
Class A
Voltage Fluctuations/Flicker
Emissions IEC 61000-3-3
Complies
Medical Electrical Equipment needs special precautions regarding EMC and needs to be installed
and put into service according to EMC information provided in this document.
Warnings: The VPAP Malibu should not be used adjacent to or stacked with other equipment.
If adjacent or stacked use is necessary, the VPAP Malibu should be observed to verify normal
operation in the configuration in which it will be used.
The use of accessories (eg, humidifiers) other than those specified in this guide is not
recommended. They may result in increased emissions or decreased immunity of the
VPAP Malibu.