General-purpose timing signal connections, Gpctr0_source signal, Figure 4-24. gpctr0_source signal timing – National Instruments E Series User Manual
Page 73: General-purpose timing signal connections -37, Gpctr0_source signal -37, Figure 4-24. gpctr0_source signal timing -37
Chapter 4 Signal Connections
National Instruments Corporation
4-37
DAQCard E Series User Manual
General-Purpose Timing Signal Connections
The general-purpose timing signals are GPCTR0_SOURCE,
GPCTR0_GATE, GPCTR0_OUT, GPCTR0_UP_DOWN,
GPCTR1_SOURCE, GPCTR1_GATE, GPCTR1_OUT,
GPCTR1_UP_DOWN, and FREQ_OUT.
GPCTR0_SOURCE Signal
Any PFI pin can externally input the GPCTR0_SOURCE signal, which
is available as an output on the PFI8/GPCTR0_SOURCE pin.
As an input, the GPCTR0_SOURCE signal is configured in the
edge-detection mode. You can select any PFI pin as the source for
GPCTR0_SOURCE and configure the polarity selection for either
rising or falling edge.
As an output, the GPCTR0_SOURCE signal reflects the actual clock
connected to general-purpose counter 0. This is true even if another
PFI is externally inputting the source clock. This signal is set to input
(High-Z) at startup.
Figure 4-24 shows the timing requirements for the GPCTR0_SOURCE
signal.
The maximum allowed frequency is 20 MHz, with a minimum pulse
width of 23 ns high or low. There is no minimum frequency limitation.
Figure 4-24.
GPCTR0_SOURCE Signal Timing
t
p
t
w
t
w
t
p
t
w
= 50 ns minimum
= 23 ns minimum