Chapter 4 Signal Connections

National Instruments Corporation

4-15

DAQCard E Series User Manual

Differential Connections for Ground-Referenced
Signal Sources

Figure 4-4 shows how to connect a ground-referenced signal source to

a channel on a DAQCard configured in DIFF input mode.

With this type of connection, the PGIA rejects both the common-mode

noise in the signal and the ground potential difference between the
signal source and the DAQCard ground, shown as V

cm

in Figure 4-4.

Figure 4-4.

Differential Input Connections for Ground-Referenced Signals

Ground-

Referenced

Signal

Source

Common-

Mode

Noise and

Ground

Potential

Input Multiplexers

AISENSE

Instrumentation

Amplifier

V

m

Measured

Voltage

AIGND

-

+

-

+

V

s

V

cm

+

-

-

+

I/O Connector

ACH<0..7>

ACH<8..15>

Selected Channel in DIFF Configuration

PGIA