Accuracy information – National Instruments 6023E User Manual

Page 84

background image

Appendix A

Specifications for PCI and PXI Buses

6023E/6024E/6025E User Manual

A-2

ni.com

Overvoltage protection

FIFO buffer size......................................512 S

Data transfers ..........................................DMA, interrupts,

programmed I/O

DMA modes ...........................................Scatter-gather

(single transfer, demand transfer)

Configuration memory size ....................512 words

Accuracy Information

Signal

Powered On

Powered Off

ACH<0..15>

±42

±35

AISENSE

±40

±25

Nominal Range (V)

Absolute Accuracy

Relative Accuracy

% of Reading

Offset

(mV)

Noise + Quantization

(mV)

Temp

Drift

(%/

°°°°C)

Absolute

Accuracy

at Full

Scale
(mV)

Resolution (mV)

Positive

FS

Negative

FS

24 Hours

1 Year

Single Pt.

Averaged

Single Pt.

Averaged

10

–10

0.0872

0.0914

6.38

3.91

0.975

0.0010

16.504

5.89

1.28

5

–5

0.0272

0.0314

3.20

1.95

0.488

0.0005

5.263

2.95

0.642

0.5

–0.5

0.0872

0.0914

0.340

0.195

0.049

0.0010

0.846

0.295

0.064

0.05

–0.05

0.0872

0.0914

0.054

0.063

0.006

0.0010

0.106

0.073

0.008

Note: Accuracies are valid for measurements following an internal E Series calibration. Averaged numbers assume dithering and averaging of
100 single-channel readings. Measurement accuracies are listed for operational temperatures within ±1 °C of internal calibration temperature
and ±10 °C of external or factory-calibration temperature. One-year calibration interval recommended. The Absolute Accuracy at Full Scale
calculations were performed for a maximum range input voltage (for example, 10 V for the ±10 V range) after one year, assuming 100 pt
averaging of data.