beautypg.com

Figure 4-25. startscan input signal timing, Figure 4-26. startscan output signal timing, Figure 4-25 – National Instruments 6023E User Manual

Page 67: Startscan input signal timing -37, Figure 4-26, Startscan output signal timing -37, Deasserted t

background image

Chapter 4

Signal Connections

© National Instruments Corporation

4-37

6023E/6024E/6025E User Manual

deasserted t

off

after the last conversion in the scan is initiated. This output is

set to high impedance at startup.

Figures 4-25 and 4-26 show the input and output timing requirements for
the STARTSCAN signal.

Figure 4-25. STARTSCAN Input Signal Timing

Figure 4-26. STARTSCAN Output Signal Timing

The CONVERT* pulses are masked off until the device generates the
STARTSCAN signal. If you are using internally generated conversions, the
first CONVERT* appears when the onboard sample interval counter
reaches zero. If you select an external CONVERT*, the first external pulse
after STARTSCAN generates a conversion. Separate the STARTSCAN
pulses by at least one scan period.

Rising-Edge

Polarity

Falling-Edge

Polarity

t

w

t

w

= 10 ns minimum

t

w

= 50-100 ns

a. Start of Scan

toff

t

off

= 10 ns minimum

Start Pulse

CONVERT*

STARTSCAN

STARTSCAN

b. Scan in Progress, Two Conversions per Scan

t

w

This manual is related to the following products: