beautypg.com

Freq_out signal, Field wiring considerations, Freq_out signal -49 – National Instruments 6023E User Manual

Page 79: Field wiring considerations -49

background image

Chapter 4

Signal Connections

© National Instruments Corporation

4-49

6023E/6024E/6025E User Manual

If you use an internal timebase clock, the gate signal cannot be
synchronized with the clock. In this case, gates applied close to a source
edge take effect either on that source edge or on the next one. This
arrangement results in an uncertainty of one source clock period with
respect to unsynchronized gating sources.

The OUT output timing parameters are referenced to the signal at the
SOURCE input or to one of the internally generated clock signals on the
devices. Figure 4-41 shows the OUT signal referenced to the rising edge of
a source signal. Any OUT signal state changes occur within 80 ns after the
rising or falling edge of the source signal.

FREQ_OUT Signal

This signal is available only as an output on the FREQ_OUT pin. The
frequency generator of the device outputs the FREQ_OUT pin. The
frequency generator is a 4-bit counter that can divide its input clock by the
numbers 1 through 16. The input clock of the frequency generator is
software-selectable from the internal 10 MHz and 100 kHz timebases. The
output polarity is software-selectable. This output is set to high impedance
at startup.

Field Wiring Considerations

Environmental noise can seriously affect the accuracy of measurements
made with your device if you do not take proper care when running
signal wires between signal sources and the device. The following
recommendations apply mainly to analog input signal routing to the device,
although they also apply to signal routing in general.

Minimize noise pickup and maximize measurement accuracy by taking the
following precautions:

Use DIFF analog input connections to reject common-mode noise.

Use individually shielded, twisted-pair wires to connect analog input
signals to the device. With this type of wire, the signals attached to the
CH+ and CH– inputs are twisted together and then covered with a
shield. You then connect this shield only at one point to the signal
source ground. This kind of connection is required for signals traveling
through areas with large magnetic fields or high electromagnetic
interference.

This manual is related to the following products: