Mode 1 input timing, Mode 1 input timing -27, Figure 4-13 – National Instruments 6023E User Manual
Page 57

Chapter 4
Signal Connections
© National Instruments Corporation
4-27
Mode 1 Input Timing
Timing specifications for an input transfer in mode 1 are shown in
Figure 4-13.
Figure 4-13. Timing Specifications for Mode 1 Input Transfer
Name
Description
Minimum
Maximum
T1
STB* Pulse Width
100
—
T2
STB* = 0 to IBF = 1
—
150
T3
Data before STB* = 1
20
—
T4
STB* = 1 to INTR = 1
—
150
T5
Data after STB* = 1
50
—
T6
RD* = 0 to INTR = 0
—
200
T7
RD* = 1 to IBF = 0
—
150
All timing values are in nanoseconds.
DATA
RD *
INTR
IBF
STB *
T1
T2
T4
T7
T6
T3