Off-board Charger/Programming Socket

You may use an off-board charger to charge the power chair batteries through the 3-pin socket located on

the front of the Europa. If you use an off-board charger, the charger current should not exceed 8 amps.

Contact your authorized Pride Provider for more information.

NOTE: The socket may also be used for reprogramming the Europa. Contact your authorized Pride

Provider for more information.

Sleep Mode (If Enabled)

The power chair controller features a sleep mode. Sleep mode is a built-in circuit that will automatically

shut off the main power if the joystick is not moved in any direction for a period of time. This time factor

is programmed into the controller. To restore power and continue, push and key on the keypad.

Europa Error Codes

The system status light is displayed within the on/off key. This light is lit if the system is turned on. It also

flashes in groups called trouble codes, to indicate system faults. The table below identifies the individual

trouble codes. If your keypad displays one of these codes, contact your authorized Pride Provider.

TROUBLE CODE
SEQUENCE

DIAGNOSIS

SOLUTION

•—•—•—

DX module fault

See your authorized Pride Provider.

••—••—••—

DX accessory fault

See your authorized Pride Provider.

•••—•••—

Left motor (or connection) fault

Check left motor wiring.

••••—••••—

Right motor (or connection) fault

Check right motor wiring.

•••••—•••••—

Left park brake fault

Check motor/brake wiring.

••••••—••••••—

Right park brake fault

Check motor/brake wiring.

•••••••—•••••••—

Low battery fault

Check that the battery wiring is secure.

•••••••—•••••••—

Over voltage fault

Check that the battery wiring is secure.

•••••••—•••••••—

CANL fault

See your authorized Pride Provider.

•••••••—•••••••—

CANA fault

See your authorized Pride Provider.

•••••••—•••••••—

Stall timeout

Turn unit on, then off.

•••••••—•••••••—

Module mismatch

See your authorized Pride Provider.

NOTE: You must turn the controller off and then on again to reset the controller, even if the source of

the fault is removed/corrected.

In the event of a fault, the system status light displays diagnostic indications. Faults are encoded as follows:

one (for a low battery voltage fault) to twelve (module mismatch) and are displayed by the light flashing the

number of times prescribed by the fault code. The flash sequence (one to twelve) is followed by a long off

period (2 seconds). If more than one fault exists, then the fault having the highest priority is indicated. The

controller must be turned off and then on again even if the source of the fault is removed. If you cannot

resolve the problem, contact your authorized Pride Provider.