17
CHECKING
NOTE:
If opening the test strip bottle for
the first time, write the discard date
(4 months from today) on the bottle
label and perform a control solution test
following the instructions below.
4.
Shake the control solution bottle. Apply 1
drop of control solution to the center of
the pink test square.
◆
The first time you open the control
solution bottle, write the discard date
on the label in the space provided. The
discard date for control solution is 3
months from the date you first open
the bottle.
◆
Check expiration date and discard
date. Do not use the control solution
after these dates.
◆
If the entire white pad surrounding the
pink test square is soaked, you have
applied too much Glucose Control
Solution.
CAUTION:
Applying too much control
solution may cause out–of–range results
and may make the meter dirty.
DCO_05074503A_OT_SS_OB_US 6/19/03 11:59 AM Page 17