Page: 65

7.3.6 DIAGNOSTIC REPORT

A Diagnostic report can be printed using diagnostic codes
0008 or 0010. Figure 7-4 is a typical report..

ID:A 11/20 15:16

S1= +0.376

BV= +2.021

BI= 0.323

SP= 0653

ST= +0.284

MV= +19.0

CS= +0.000

S2= +1.20

HV= +0.920

HI= +0.975

SP= 400

ST= +0.009

MV= +19.4

CS= +0.000

Figure 7-4

Typical Diagnostic Report

Table 7-2 is a description of the information contained in
a typical diagnostic report

Report

Description

ID:A

Controller Identification

11/20 15:16

Time report was generated

S1

+0.376

Data that follows is for Module 1
Sensor Signal is +0.376 VDC

BV

+2.021

Bridge Voltage 2.021 VDC

BI

+0.322

Bridge Current 323 mA

SP

0653

Sensor Power 653 mW

ST

+0.284

Sensor Type
(Where: .2 -.3 VDC: Combustible)

MV

+19.0

Module Voltage = 19 VDC

CS

+0.000

Calibration Switch
Off 0.000, On = 2.5 VDC

S2

+1.20

Data that follows is for Module
Sensor Signal is + 1.20 VDC

HV

+0.920

Heater Voltage Signal

HI

+0.975

Heater Current Signal

SP

400

Sensor Power 400 mW

ST

+0.009

Module Type
(Where: .11-.1 VDC: H2S)

Table 7-2

Diagnostic Report Description