Loc - go to local – SIGMA LBA-708 User Manual
Page 253

Key Type Value
Description
Operator’s Manual
LBA-PC
Doc. No. 10654-001, Rev 4.10
253
then local control is restored (equivalent to password lockout being disabled).
5 = 32x
EnergyOfBeam
4
F
calibrated energy of beam entered in
Computations dialog
EnergyOfFrame
4
F
raw frame total equivalent to
EnergyOfBeam
EnergyUnits
4
L
0 = j
1 = mj
2 = uj
3 = nj
u = pj
5 = w
6 = mw
7 = uw
8 = nw
9 = pw
10 = fl
AC
4
L
-1 = ultracal was on but capture location,
frame size, or resolution did not match
0 = ultracal off
>0 = ultracal subtracted
RS
4
L
-1 = reference subtract was on but capture
location, frame size, or resolution did not
match
0 = reference sub off
1 = reference sub subtracted
GC
4
L
-1 = gain correct was on but capture
location, frame size, or resolution did not
match
0 = gain correct off
1 = gain correct multiplied
CommentLine
3
S
up to 256 characters
WriteProtect
3
B
1 = cannot overwrite this frame
A.5.4.8 LOC - go to local
When the LOC message is sent or when the LBA-PC goes from remote to local via the REN line