beautypg.com

Pfs? - pass/fail status, Coordinate systems, Spatial coordinates – SIGMA LBA-708 User Manual

Page 200: 7 coordinate systems

background image

Operator’s Manual

LBA-PC

200

color or gray scale and displayed in the beam window.

FrameNumber=52;

CommentLine=This will appear in the title bar;

WriteProtect=1(^END)

10.6.3

PFS? - Pass/Fail Status

To retrieve the current pass/fail status, you must send the PFS? command.
The LBA-PC responds by repeating PFS followed by a “key=value” parameter for each value that has
pass/fail testing enabled. The ‘key’ is an ASCII text string identical to the label displayed in the left-

hand column of the results window. The ‘value’ is a boolean, where 0 equals fail and 1 equals pass.
The following example describes how the host controller retrieves the current pass/fail testing status.

Total, Peak, Centroid X, and Centroid Y are being tested. Total passed, Peak failed, Centroid X failed,

Centroid Y passed.
Host sends

:PFS?


LBA-PC sends

PFS

Total=1;
Peak=0;
Centroid

X=0,

Centroid=1(^END)

10.7 COORDINATE SYSTEMS

10.7.1 Spatial Coordinates

The visible spatial coordinate system of the LBA-PC is called world coordinates. World coordinates are

used for computed centroid location and peak location, aperture location, cursor and crosshair
location, and pass/fail centroid and peak locations. World coordinates are defined by the Origin

Location, ‘Lens’ setting, and Pan Location. The world coordinates system is the only coordinate

system that is used when using the LBA-PC in the local, non-remote mode.
When operating the LBA-PC remotely, you must be aware of two other coordinate systems called

detector and frame coordinates. Detector coordinates are represented on the LBA-PC display by the

Pan/Zoom window. Detector coordinates are used to describe the active pixels of the camera
detector. Frame coordinates are used to describe the frame data stored in the frame buffer.
The figure below shows the relationship between the Pan/Zoom window, the capture window, frame

data, and the Beam display window. The Pan/Zoom window represents the active area of the camera

detector. The capture window, represented in the LBA-PC display by the dark gray square inside the

Pan/Zoom window, represents the location and size of the area of the detector that will be digitized.

The digitized data is placed in the frame buffer. Data in the frame buffer is then converted to false

This manual is related to the following products: