Series 9205 Operations Manual Rev A

79

Chapter 3 – Configurator 2.0 Menus

The following section will detail the menus that are found with the Configurator 2.0 software for the 9205

controller.

Burnoff

Configurator - Burnoff menu option

When a probe is in a furnace, soot will collect in the end of the probe, which will have a negative effect on

the performance of the probe. Burnoffs are used to clean out the built-up carbon by burning it off of the

probe.

Burnoff:

Clicking on this value allows users to manually initiate burnoff procedure. The user will have to confirm

this action

Configurator – Confirm burnoff

The “Test Status” will show that a burnoff is in progress.

Impedance Test:

Clicking on this value manually turns on/off the impedance test (this can

only

be initiated manually). This

test lasts 30 seconds. The user will have to confirm the action.