beautypg.com

Chapter 4 – opcodes – Super Systems 9205 Series User Manual

Page 154

background image

Series 9205 Operations Manual Rev A

153

Chapter 4 – OPCODES

Temperature Value Column

Atmosphere Value Column

Option Value Column

AB_SOAK

N/A

N/A

Soak Time

This sets a soak time for an autoboost step in a recipe. It must be monitored externally by SSi's realtime

carbon diffusion modeling software which will adjust the soak times based on a predefined target model for

the specified work.

ABD_Model

N/A

N/A

Model

This opcode instructs the Real Time Carbon Diffusion (RTCD) program what model to use when the

autoboost or autodiffuse is encountered.

AD_SOAK

N/A

N/A

Soak Time

This sets a soak time for an autodiffuse step in a recipe. It must be monitored externally by SSi's realtime

carbon diffusion modeling software which will adjust the soak times based on a predefined target model for

the specified work.

ALARM

N/A

N/A

User Alarm Number

The alarm function is used to notify the operator that an operation is complete or that a manual action is

required. By use of the

Settings

menu screen, up to 99 User Alarms can be assigned, with a short text on

each that is displayed during an active alarm condition. The program waits until the alarm is acknowledged

to proceed. User Alarm 0 is used to turn off a user alarm. Its function is the same as the acknowledge

button

.

ATM_INQ

N/A

ATM Level

Wait, Wait Up, Wait Down

The atmosphere inquiry is used to wait for the actual atmosphere to reach the set point. The default band

can be set under the Configuration Menu and is typically 10 (i.e. 0.10 percent carbon).

The

SET_WAIT

opcode will change the band limit

The

LIMIT

opcode immediately following this opcode sets a time limit on the wait.

A

BRANCH

opcode immediately following this opcode can be used to change the program flow

based on the inquiry results.

BO_start

N/A

N/A

Start, start+wait, check+wait

This opcode will allow the start of a burnoff from the controller.

BRANCH

True Step Number

False Step Number

N/A

The Branch opcode can change program flow based upon an inquiry opcode. The True Step Number is the

program step to go to if the inquiry is evaluated as true; and the False Step Number is the program step to

go to if the inquiry is evaluated as false.

The True Step Number ranges from 1 – 24.

The False Step Number ranges from 1 – 24.

CC_SP_L

Setpoint Lower Limit

Setpoint Upper Limit

N/A

This manual is related to the following products: