SPECIFICATIONS
STANDARD FEATURES
SCALE SPECIFICATIONS
Display
Bright red LED; 1-inch x 6-digit numeric
Status Indicators
Gross, Net, Motion, Zero, LB, Kg (LED)
Keyboard
19-key sealed tactile feel membrane
Internal Resolution
20-bit A/D Sigma-Delta; 1,048,000 d
Display Resolution
120,000 dd industrial; 10,000 dd HB44
Display increments
Selectable 1, 2, 5, 10,20, 50 100
Decimal Point
Selectable 0, 1, 2, 3, 4 decimal places
Conversation Rate
60 samples/second typical
Signal Sensitivity
0.1 uV/graduation (min)
Signal Range
0.5 mV/V to 6mV/V
Load Cell Excitation
10
±
0.5VDC (-5 to +5)
Load Cell Power
12 x 350
Ω
or 24 x 700
Ω
load cells
Auto Zero Tracking
0-60 dd in 1/4 dd increments
Auto Zero Delay
0-25 seconds In 0.1 second increments
Motion Detect
0-60 dd in 1/4 dd increments
Motion Delay
0-25 seconds In 0.1 second increments
Digital Filter
0-18 selectable filter (DSP) levels
Calibration
Selectable multi-point (up to 5) digital
calibration to linearize input signal
Watchdog Timer
Enable/disable fault tolerant operation
RFI Protection
Signal, excitation and sense lines
RAM
32K provides 500 ID storage (parts, etc)
COMMUNICATIONS
Serial Port 1
Simplex RS232 or 20mA current loop
Serial Port 2
Full duplex RS232 or RS485
Digital Port
3 inputs and 3 outputs; Low active (TTL)
Print Formats
4 user configurable print formats
GENERAL SPECIFICATIONS
Power
110/220VAC ±10% @ 50/60Hz
or 12-28 VDC
Power Consumption
10W @ 115VAC; 6W @ 12VDC
Operating Temperature -10 to +40
o
C.
Weight
6 lbs (2.72 kg)
Warranty
One year limited
OPTIONS
Analog Output Module
Fully isolated 0-10VDC or 4-20mA; 16-bit
resolution; 650
Ω
load resistance
Relay Board
4-channel external relay enclosure
Relay Modules
Input/output solid-state (AC/DC) relays
Clock Module
Time & Date clock Y2K and Leap Year
High speed (HS)
100 samples/second conversion rate;
High speed process control applications
EZ-LINK
PC software; Facilitates configuration
and Scale Basic programming
IS Barriers
Hazard location intrinsic safety barriers
MODES OF OPERATION
The IDS 430/433 is a versatile indicator for use in various industries
and applications that can range from basic weighing, material testing
to more complex formula batch weighing. The IDS 430/433 is
housed in a rugged stainless steel enclosure and provides a
universal mounting bracket for wall or desk mounting. The viewing
angle can be adjusted for the best possible view of the weight. The
bright red 1 inch high LED display makes it easy to read the weight
from longer distances compared to other indicators. The keyboard
features large keys, which give you tactile and audible confirmation
of data entry. The IDS 430/433 is programmable in an easy and
flexible macro language called Scale Basic
. In addition to the built-
in modes of operation, Scale Basic allows you to customize the
operation of the IDS 430/433 to meet your application requirements.
The Scale Basic language provides various commands and
functions that include: math operations, I/O control, setpoints,
timers, data entry, id storage, message display, and program
sequence control among others. Programming the IDS 430/433 can
be performed through the keyboard or a PC using the optional EZ-
LINK
software that greatly facilitates setup and programming.
MODE 0 - Normal
Basic gross, tare and net weighing
MODE 1 – Fill to Setpoint
Single setpoint fill with relay output control and weight print
MODE 2 - Over/Under Checkweigh
Provides 3 TTL outputs and 2 setpoints to indicate over,
under and between weight conditions. Outputs can activate
lights or alarms
MODE 3 – Vehicle Weigh-In/Out with Totals
Weigh-In/Out program for up to 500 vehicles, maintains
running totals for each vehicle and provides reporting
functions
MODE 4 - Multi-Container/Axle Auto Weighing
Auto weighs and prints each weight placed on the scale, each
weight is sequentially numbered and accumulated, the total
can be printed manually or automatically and provides traffic
light control
MODE 5 - Auto Axle Weigh (Long Scales)
Same as Mode 4 but previous axles are auto tared
MODE 6 – ID Tare, Print and Total
Provides tare entry and recall by ID, prints ID, sequence #
and GTN weights, keeps running totals by ID, provides
reporting functions
MODE 7 - Peak Hold Mode
Display or print the peak gross or net weight detected
A = 7.63” (19.38 cm)
B = 6.88” (17.48 cm)
C = 8.56” (21.74 cm)
D = 2.00” (5.08 cm)
E = 3.00” (7.62 cm)
D
E
12345
A
B
C