IDS 430/433 Users Manual Version 1.0
- 55 -
Function 74 Configure Timers
The display prompts “tr” when function 74 is activated. Enter the timer number to
be modified (1 to 5) then press the ENTER key. The first parameter (L) is the
time length in tenths of a second. The second parameter (Fn) is the function
number to be executed when the timer times out.
Parameter 75 Enable Peak Detect
Enter a 1 to enable peak detect, enter a 0 to disable peak detect.
Parameter 76 Debug mode for Event monitors and Scale Basic
The Debug mode is used to ‘debug’ event monitors and scale basic functions.
Enter a 1 to activate debug mode, enter a 0 to disable debug mode. When debug
mode is enabled Scale Basic functions ‘single step’. The subroutine stack level
and the Function number are displayed when entering a function. Example: “0
F12”. Press the ENTER key to begin execution of the function. The instruction
address and instruction code are displayed. Example: “21 255”. Continue to
press the ENTER key to step thru the program. NOTE: The first step displayed
for all scale basic functions is a GOTO (244).
Function 78 Initialize ID memory
Use function 78 to initialize id memory after it has been corrupted. The ID
memory becomes corrupted when power is lost (when battery backup is
disabled) or when the battery is allowed to discharge below 2 volts.
Parameter 79 Enable Multi-point Linearization
Enter a to 1 to enable multi-point Linearization, enter a to 0 to disable multi-point
Linearization. The IDS 430/433 uses the calibration factors set in function 80
below when multi-point Linearization is enabled.