beautypg.com

Industrial Data Systems IDS 430 User Manual

Page 58

background image

IDS 430/433 Users Manual Version 1.0

- 58 -

Diagnostic Test 2: Serial Com Port 1 – Display Errors

This test displays the number of framing and parity errors that are detected in the

input data stream. The error count is set to zero when entering this test. New

errors are displayed as “PxxFxx” where Pxx is the number of parity errors and

Fxx is the number of framing errors. If the error count exceeds 99, the display will

remain at 99. Press the CLEAR key to exit Diagnostic test 2.

Diagnostic Test 3: Serial Com Port 1 - Transmit Data

This test transmits data out of serial communications port 1. The display

prompts: “OUT 0”. Press the ENTER key to send data. Serial Port 1 transmits

“1234567890ABCDEF” and prompts “OUT 1” to indicate that 1 transmission has

occurred. Press the ENTER key again or press the CLEAR key to exit diagnostic

test 3.
NOTE: If the error message “Er CtS” appears, the data cannot be transmitted

because the handshake signal, clear-to-send (TB1 pin 7), is low or not

connected. If clear-to-send is not used jumper it to data-terminal-ready (TB1 pin

4)

Diagnostic Test 4: Serial Com Port 2 - Display Input Data

This test displays serial data as it is received by serial communications port 2.

The numeric display has limited alpha display capability, however numeric and

some alpha characters are legible. If all data being received is unintelligible, the

baud rate is probably incorrect. If nothing is displayed on the display, then check

the following:

1

. RS232/RS485 jumper on the main circuit board.

2. Cabling between the communicating units.

3. Make sure the sending unit is actually sending data.

Press the CLEAR key to exit diagnostic test 4.

Diagnostic Test 5: Serial Com Port 2 – Display Errors

This test displays the number of framing and parity errors that are detected in the

input data stream. The error count is set to zero when entering this test. New

errors are displayed as “PxxFxx” where Pxx is the number of parity errors and

Fxx is the number of framing errors. If the error count exceeds 99, the display will

remain at 99. Press the CLEAR key to exit Diagnostic test 5.

Diagnostic Test 6: Serial Com Port 2 - Transmit Data

This test transmits data out of serial communications port 2. The display prompts:

“OUT 0”. Press the ENTER key to send data. Serial Port 2 transmits

“1234567890ABCDEF” and prompts “OUT 1” to indicate that 1 transmission has

occurred. Press the ENTER key again or press the CLEAR key to exit diagnostic

test 6.

This manual is related to the following products: