beautypg.com

8 installing a new no sensor – LumaSense Technologies INNOVA 1316A-3 User Manual

Page 82

background image

Chapter 4
_________________________________________________________________________________________

_____________________________________________________________________________

BE6024-12

1316A-3 Multi Gas Monitor

LumaSense Technologies A/S

Instruction Manual

Page 82 of 96

Figure 4.34. Status window showing a CO2 span failed.

If the NO channel is being span calibrated, the current full-scale NO

output is calculated and saved. If the current full-scale NO output is less

than or equal to a preset value (The Low NO Sensor Limit) the New NO

Sensor Required is shown in the Status window, see

figure 4.13

.

After making the gain Calibration please make sure that the Mode is set

back into 5 and that the Zero calibration is set to “Auto”, as described in

section 4.3.5

.

4.3.8 Installing a new NO sensor.


After installation the new NO sensor it must be span calibrated following

the procedure described in the previous sections.
After performing the NO span calibration:
Select the New NO sensor in the Bench command menu, see figure

4.35.