Figure 4.34. Status window showing a CO2 span failed.
If the NO channel is being span calibrated, the current full-scale NO
output is calculated and saved. If the current full-scale NO output is less
than or equal to a preset value (The Low NO Sensor Limit) the New NO
Sensor Required is shown in the Status window, see
After making the gain Calibration please make sure that the Mode is set
back into 5 and that the Zero calibration is set to “Auto”, as described in
4.3.8 Installing a new NO sensor.
After installation the new NO sensor it must be span calibrated following
the procedure described in the previous sections.
After performing the NO span calibration:
Select the New NO sensor in the Bench command menu, see figure
4.35.