6 operating modes of the multi gas monitor 1316a-3, 1 start-up operating mode – LumaSense Technologies INNOVA 1316A-3 User Manual
Page 49
Chapter 3
_________________________________________________________________________________________
_____________________________________________________________________________
BE6024-12
1316A-3 Multi Gas Monitor
LumaSense Technologies A/S
Instruction Manual
Page 49 of 96
Figure 3.11: The Soft Reset Window.
Press the Send Button and the Instrument is Reset and ready for set-
up.
3.6 Operating Modes of the Multi Gas Monitor
1316A-3
The Multi Gas Monitor 1316A-3’s operations comprise four modes of
operation.
¨
Start-up
¨
Normal
¨
Standby
¨
System Fault
3.6.1 Start-up Operating Mode
After POR (power on / reset), the Multi Gas Monitor 1316A-3
performs the following:
¨
Start-up and initialization.
¨
Self-test.
¨
Thermal warm-up. After a cold start or a Software Reset
bench command the warm-up time period is 35 seconds
maximum. After leaving standby mode the typical warm-