4 the ndir temperature measurements – LumaSense Technologies INNOVA 1316A-3 User Manual
Page 14
Chapter 1
_________________________________________________________________________________________
_____________________________________________________________________________
BE6024-12
1316A-3 Multi Gas Monitor
LumaSense Technologies A/S
Instruction Manual
Page 14 of 96
µm (2260 cm
-1
), and the filter associated with Carbon Monoxide is
centred on 4.65 µm (2150 cm
-1
).
In the measurement chamber the three beams, which differ in
wavelength frequency, each excite their respective gas (if present).
The absorption of the incident light will be registered by the
Detector.
System Board Assembly
The system board assembly is a surface-mount printed circuit board
that includes the digital and analogue electronic components and the
IR detector assembly. The IR detector assembly includes thermopile
IR detectors for the Ethanol, CO, CO
2
, and reference channels. Light
passed through the band pass filters is directed through the
thermopile window to a detector substrate. The substrate has a light-
sensitive coating, which produces a voltage that is proportional to
light intensity. A reference voltage is produced by the reference
detector, which provides an output that varies with light intensity
and is not affected by the sample gas. The Ethanol detector includes
a thermistor. The detector block temperature measurement
facilitates compensation for IR filter and IR detector temperature
variances. The detector outputs are amplified and sent through a
multiplexer to an ADC converter. The microprocessor samples the
ADC converter output at 200 Hz and derives an uncompensated
digital value representing each gas concentration. The data are
subjected to a compensation routine that provides adjustment for
temperature and pressure variances. The digital value representing
each gas concentration is derived.
1.4.4 The NDIR Temperature Measurements.
The analyzer incorporates several thermistors for the measurement
of temperatures.
IR Detector/Filter Temperature (Ethanol Temperature):
A thermistor located in the optical assembly measures the IR
detector and IR filter temperature. It is used as follows:
“Gas law compensation” and detector response compensation during
Ethanol, CO and CO
2
gas concentration measurement.
Span calibration (via the use of temperature compensated Ethanol,
CO
2
and CO data).
An “out of range” IR detector/filter temperature condition causes
Ethanol, CO
2
and CO data to be declared “data invalid.”