10.4 Specifications
TRANSDUCER INPUT
Transducer type
full-bridge, strain gage
Excitation Voltage
5 or 10 VDC,
short circuit protected with sensing
Transducer full-scale output
.5 to 11.0 mV/V @ 5V excitation
.5 to 5.5 mV/V @ 10 V excitation
Amplifier Gain Selection
automatic
Calibration Type
shunt, mV/V or
2-, 3- or 5- point known load
Differential Input Voltage
+/- 55mV (max.)
A/D Converter
24-bit Sigma-Delta
Low-pass filter
digital, 24-tap FIR
Resolution and Frequency Response
see “FREQ. RESPONSE Menu Item” on
INSTRUMENT-ONLY ACCURACY
(Sense wires used;Frequency Response setting<=16Hz;
Linearity, repeatability & hysteresis)
mV/V Calibration
±0.1% F.S.
Known -Load Calibration
±0.01% F.S.
AUXILLIARY INPUTS
Quantity
2
Type
momentary contact closure
Response Time
< 5ms
Field-Selectable Functions
tare on, tare off,
peak/valley clear, peak/valley hold,
track hold
ANALOG OUTPUT
Output voltage range
5, ±5, 10 or ±10 VDC (field selectable)
Output current range (optional current
output channels)
4-20 mA
Source
any channel’s track, peak or valley value
Isolation
500V
Resolution
13 bits
Frequency Response
same as input when driven by the same
channel’s tracking data