10.2 Wiring

Connect the transducer to a Strain-Gage Input channel by wiring it to the 12-pin con-
nector of that channel. The Customer Information Sheet that shipped with the instru-
ment describes which cards are installed in each channel. The pin-out for this
connector is shown on the following table.

The Analog Output and Analog Return pins are electrically isolated from all other pins
on the instrument.

The maximum recommended cable length to a Signature Module is 100 feet [30m].

Figure 10-1: Unamplified Transducer Connection to Strain-Gage Input Channel

Table 4-3: Strain-Gage Input Channel Pin Connections

Pin

Label

Function

Reference Pin

1 (top)

+EXC

(+)Excitation

10

2

+SEN

(+)Sense

10

3

-SEN

(-)Sense

10

4

-EXC

(-)Excitation

10

5

+SIG

(+)Signal

10

6

-SIG

(-)Signal

10

7

+OUT

Analog Output

8

8

-OUT

Analog Return

-

9

+MEM

(+)Signature

10

10

-MEM

(-)Signature / Digital Ground

-

11

AUX1

Auxiliary Function 1

(connect to pin 10 to activate)

10

12

(bottom)

AUX2

Auxiliary Function 2

(connect to pin 10 to activate)

10

NOTICE

If you are not using remote sensing of excitation, jumper (+)Sense to (+)Excita-
tion and (-)Sense to (-)Excitation. If nothing is connected to (+)Sense and (-
)Sense, the instrument will enter the ERROR mode with error code 46.