Honeywell SC1000 User Manual
Page 118
page 118 008-0608-00
If the
CALIBRATION TYPE
is 5-Point Known Load Calibration...
•
The display will read
DOING 5POINT CAL
, and prompt you to
APPLY 00000 UNIT
(where 00000 and UNIT are previously entered as
KNOWN POINT 1/5
and
DISPLAY.
UNITS
). When you have applied this load to the transducer, press [ENTER].
•
The display will read
WORKING
, then
APPLY 00250 UNIT
(where 00250 and UNIT are
previously entered as
KNOWN POINT 2/5
and
DISPLAY. UNITS
). When you have
applied this load to the transducer, press [ENTER].
•
The display will read
WORKING
, then
APPLY 00500 UNIT
(where 00500 and UNIT are
previously entered as
KNOWN POINT 3/5
and
DISPLAY. UNITS
). When you have
applied this load to the transducer, press [ENTER].
•
The display will read
WORKING
, then
APPLY 00750 UNIT
(where 00750 and UNIT are
previously entered as
KNOWN POINT 4/5
and
DISPLAY. UNITS
). When you have
applied this load to the transducer, press [ENTER].
•
The display will read
WORKING
, then
APPLY 01000 UNIT
(where 01000 and UNIT are
previously entered as
KNOWN POINT 5/5
and
DISPLAY. UNITS
). When you have
applied this load to the transducer, press [ENTER].
•
The display will now read
RETURN PNTS.=NO
. If you press [ENTER], the display
will indicate
DONE
and the instrument will return to the RUN mode. If you use
the [UP] button and [ENTER] to select
RETURN POINTS=YES
, you may now apply
the same loads again but in reverse order. This will allow the instrument to
measure and partially eliminate any hysteresis that may be present in the
transducer. After the loads have been applied again, the display will indicate
DONE
and the instrument will return to the RUN mode.