Table b-6. error messages – KEPCO TMA 4882-27 User Manual
Page 65

TMA4882-27 021910
B-13/(B-14 Blank)
TABLE B-6. ERROR MESSAGES
ERROR MESSAGE
EXPLANATION
0,“No error”
No error
-100,“Command error”
Command and data understood, but more information included which is not
recognized.
-102,”Syntax error”
First 4 characters recognized, subsequent characters not recognized.
-103,”Invalid separator”
For example, VOLT.10 received instead of VOLT:10
-108,”Parameter Not Allowed Error”
Volt12 sequence, channel number is invalid
-109,”Missing parameter”
For example, VOLT instead of VOLT 21.
-111,”Header separator error”
Missing space between volt and value or ; missing
-113,”Undefined header”
First 4 characters could not be identified as legal command.For example,
command VLT instead of VOLT
-120,”Numeric data error”
Expected number but other characters were detected
-121,”Invalid character in number”
Volt 1,500 (comma not allowed)
-123,”Exponent too large”
Exponent E+3 or greater is invalid.
-141,”Invalid character data”
For example OUTP OFD or OUTP STOP instead of OUTP OFF
-150,”String data error”
Invalid characters were detected in numeric entry.For example E.1 instead
of E+1 or 4d3 instead of 4.3.
-222,“Data out of range”
Value exceeds power supply rating
-223,”Data format error”
Multiple decimals in digit, Multiple E, etc.
-224,“Illegal parameter value”
For example, OUTP 2 instead of OUTP 1
-241,”Hardware missing”
Requesting device 2 status (INST:NSEL 2)
-350,”Queue overflow”
More than 15 errors are in queue.
-410,”Query interrupted”
New command sent before data from previous query read. Previous query
data lost.
-430,”Query Deadlocked“
Over 255 characters received in single input string”