beautypg.com

Power up tests – Rose Electronics Porter User Manual

Page 18

background image

Power Up Tests

14

PORTER INSTALLATION AND OPERATIONS MANUAL

Power Up Initial Display

Push in the power switch to turn the unit on. Porter goes through a power up self-test

which checks the main functions of its electronics. The leftmost LED labeled POWER

should glow green and the other red LEDs should all turn on. Then the unit will cycle

through four tests lighting up LEDs 1 through 4 as each test is performed. Errors are

indicated by the ERROR LED lighting and the unit halting. See TABLE II for the full

sequence of the power up LED display and TABLE III for error displays.

Table II LED Power up Display

Legend: M=Mode LED D=Data LED B=Busy LED E=Error LED

= LED off

= LED on

M D B E

1

2

3

4

5

6 7

8

1st

All LEDS on

2nd

Test 1: EPROM checked

3rd

Test 2: 16K RAM buffer

Test 2: 64k RAM buffer

Test 2: 256k RAM buffer

4th

Test 3: Static ram

5th

Test 4: Non-Volatile ram

6th

Unit ready

Program Checksum Test

This test verifies that the unit’s internal program is valid by performing a checksum test.

If a fault exists, a program checksum error shows on the LEDs as error 1, as shown in

TABLE III. The unit will immediately halt. This error requires that the unit be serviced.

Buffer Memory Sizing And Read/Write Test

In this test, the Porter automatically determines the amount of butter memory present

and indicates the amount via the status LEDs as shown in TABLE III. The buffer memory

is then tested to ensure that the memory system is operational. Any failure will cause the

error 2 condition to be displayed, as shown in TABLE II, indicating that a buffer memory

data error has occurred. The unit will not continue its further tests and will halt. An error

indicates that the unit requires servicing.

Static Ram Read/Write Test

The next test is for static ram. Since the Porter has no static ram, this test is displayed as

shown in TABLE II for consistency with other Rose products.

Non-volatile Ram Tests

The fourth and final power-up test is for the non-volatile memory. A series of tests are

performed. An error shown by the LEDs as error 4 indicates that the data in the non-

volatile memory has been corrupted. To attempt to rewrite the default parameters push

the SELECT switch then the ADVANCE switch. If the error condition goes out, the unit

may be powered up again to see if it halts at error 4 again. If it still displays error 4, a fatal

error has occurred and servicing is required. If the unit displays an error 45, the default

parameters were not written correctly. This is a fatal error and servicing is required.

If the unit recovers from the error the configuration parameters have been returned to

their default settings and the unit must be configured again if these default settings are

unacceptable. This condition should not normally occur.