Round horizontal tank calibration points, Number of points: 10 maximum error: 0.3% fs, Number of points: 8 maximum error: 0.5% fs – Precision Digital PD650 User Manual
Page 40: Calibration error ( error)
Model PD650 Large Display Process Meter
Instruction Manual
40
Round Horizontal Tank Calibration Points
Number of Points: 10 Maximum Error: 0.3% FS
Input mA
Display
(%
Volume)
1 4.00
0.00
2 4.80
1.80
3
6.00
7.20
4 7.20
14.20
5 9.20
28.10
6 14.80
71.80
7 16.80
85.80
8 18.00
92.80
9 19.20
98.10
10 20.00
100.00
Number of Points: 8 Maximum Error: 0.5% FS
Input mA
Display
(%
Volume)
1 4.00
0.00
2 4.80
1.90
3 6.40
9.50
4 8.40
22.40
5 15.60
77.70
6 17.60
90.50
7 19.20
98.10
8 20.00
100.00
Calibration Error (error)
A meter display of Error during calibration indicates that the calibration process
was not successful. The meter should be recalibrated.
The Error message will appear if input 1 signal and input 2 signal are too close
together. Refer to Minimum Input Span (Error Message), page 35.
The Error message will appear if input 1 signal is inadvertently also applied for
input 2 calibration, or ENTER is pressed before applying input 2.