Precision Digital PD650 User Manual
Page 14

Model PD650 Large Display Process Meter
Instruction Manual
14
Isolated 4-20 mA Transmitter Output
CALIBRATION 
RANGE 
The transmitter output can be calibrated so that a 4 mA output 
is produced for any process/rate measured by the meter. The 
20 mA output may correspond to any process/rate that is at 
least 501 counts greater or smaller than the process/rate 
corresponding to 4 mA. (Ex. 4 mA = 0, 20 mA = 501) If the 
span between 4 and 20 mA is less than 501 counts, an 
error
message will appear.
EQUIPMENT 
NEEDED 
The 4-20 mA output from the meter is calibrated without the 
use of a calibrator. 
OUTPUT LOOP-
POWER 
24 VDC ± 5% @ 20 mA, regulated. Maximum loop resistance 
is 1200 
Ω. Isolated from input loop-power.
ACCURACY
± 0.1% FS ± 0.004 mA
ISOLATION
500 V input-to-output or input/output-to-24 VDC supply
EXTERNAL 
LOOP-POWER 
SUPPLY 
35 VDC maximum
OUTPUT LOOP 
RESISTANCE 
Power supply
Minimum loop 
resistance 
Maximum loop 
resistance 
24 VDC
10
Ω 600
Ω
35 VDC (external)
600
Ω 1000
Ω