Inp 1, Inp 2, Dis 1 – Precision Digital PD6200 User Manual
Page 43: Dis 2

Model PD6200 Analog Input Rate/Totalizer
Instruction Manual
43
Calibrating the Meter with External Source (Cal)
To scale the meter without a signal source
refer to Scaling the Meter (SCALE) page 39.
The meter can be calibrated to display the process variable in engineer-
ing units by applying the appropriate input signal and following the cali-
bration procedure. 
The use of a calibrated signal source is strongly recommended to cali-
brate the meter. 
 
 CAL
 
Prog
 Inp 1
 
CAL
 Inp 1
 
I n p 1
 Inp 2
 
CAL
 Inp 2
 
I n p 2
 dis 1
 
Inp 1
00
0
.000
d i s 1
 dis 2
 
Inp 2
1
00.000
d i s 2
T he display
flashes while
sensing the
inputs
Warm up the meter for at least 15 minutes
before performing calibration to ensure
specified accuracy.
                    This manual is related to the following products:                                    
                                                                        