Pressure Systems, Inc.

NetScanner™ System (9016, 9021, & 9022) User’s Manual

www.PressureSystems.com

3 - 23

Command ‘C’— Sub-command Index 03: Abort Multi-Point Calibration

This sub-command has no additional parameters.

Command

C 03

‘C’ is the command letter.
‘ 03’ is the sub-command index (ii) for Abort.
NOTE: all parameters are separated by a space.

Response

A

‘A’ is the acknowledge letter

Description: Aborts the Multi-Point Calibration function, if it is currently in progress. This

sub-command also restores the original “A/D samples for averaging” count to the
module that was in use before the calibration function was started.

It should be noted that executing the Configure & Start (‘00’) sub-command again,
after the calibration function has started collecting data (per Collect Data (‘01’) sub-
commands), but before the final data are calculated (per Calculate & Apply (‘02’)
sub-command), will have the same affect as this Abort function.

Example:

!

Abort the Multi-Point Calibration function previously started

C 03

Read response:

“A”