6. Internal damping filters out noise or the effects of rapid changes in the variable without

affecting measurement accuracy.

The damping value is the filter coefficient that approximates the time required for the output to
achieve 63% of its new value in response to a step change at the input. The actual time depends
on many factors, including sensor type and density of the process fluid. The meter rounds
down the chosen damping value to the nearest programmed filter coefficient. Programmed
filter coefficients, in seconds, are:

If desired, enter an internal damping value into the Damping text box in the Flow panel.

Table 7-2

Effect of flow direction on outputs and totalizers

Fluid flow
direction

Output or
totalizer

Flowmeter configuration

Forward

Reverse

Fluid flowing
in same direction
as flow arrow
on sensor

Milliamp output

Output increases as flow rate
increases

Output goes to 3.8 mA

Pulse output

Output increases as flow rate
increases

Output remains at 0 Hz

Internal totalizers

Flow totals increase

Flow totals remain constant

Fluid flowing
in opposite direction
from flow arrow
on sensor

Milliamp output

Output goes to 3.8 mA

Output increases as flow rate
increases

Pulse output

Output remains at 0 Hz

Output increases as flow rate
increases

Internal totalizers

Flow totals remain constant

Flow totals increase

0.1

0.2

0.4

0.8

1.6

3.2

6.4

12.8