Command 4 – conversion equation setup, Command 5 - data range setup – Casio EA-200 User Manual
Page 50
20020601
English
Command 4 – Conversion Equation Setup
{ 4, Equation Number, Equation Type, Number Format, Constants }
Equation Number
Equation Type
Number Format
Constants
* 0
Clear All equations.
* 0
Clear equation selected by the equation number parameter.
–––
–––
1
Equation 1 (Channel 1)
1
Polynomial
K0 + K1X + K2X
2
+...+ KnX
n
*
1
* 0
Standard
K0( ,K1,...,K9) *
3,4
2
Equation 2 (Channel 2)
2
Mixed polynomial
KmX
–m
+ ...+ K–1X
–1
+ KO + K1X +...+ KnX
n
*
2
10
Integer part (Decimal part cut off.)
K–4( ,..., K–1, K0, K1, ..., K5) *
3,4
3
Equation 3 (Channel 3)
3
Power
K0 • X
(K1)
+ K2
4
Modified power
K0 • K1
X
+ K2
5
Logarithmic
K0 + K1 In(X)
6
Modified logarithmic
K0 + K1 In(1 / X)
7
Exponential
K0e
(K1X)
+ K2
K0( , K1, K2, K3) *
3,4
8
Modified Exponential K0e
(K1/X)
+ K2
9
Geometric
K0X
(K1X)
+ K2
10
Modified geometric
K0X
(K1/X)
+ K2
11
Reciprocal logarithmic 1 / {KO + K1 In(K2X)} + K3
12
Steinhart-Hart
1 / {KO + K1(In 1000X) + K2(In 1000X)
3
} + K3
4
Equation 4 (SONIC channel)
* 0
Clear equation 4
Unit
Temperature
1
Temperature used by distance conversion expression
* 0
°C(Celsius)
Temperature
1
°F (Fahrenheit)
( *20)
2
°C (Celsius)
3
K (kelvin)
4
°R (Rankin)
α-1-3
Channel Select
Data Select
Data Begin
Data End
Step
K
FFT Samples
* 0
Current send channel
* 0
Raw data
1 to 120000
1 to 120000
Data Range Steps
(*255)
1 to 14 : Samples used
1
Channel 1
1
d/dt
(*1)
* 0 : Last sample
–1: Data range number / K
(*6) 2
n
(2–16384)
2
Channel 2
2
d
2
/dt
2
(*1)
3
Channel 3
10
FFT-Real
4
SONIC channel
11
FFT-Imaginary
5
DIG IN channel
6
Recorded time data
10
Microphone
Command 5 - Data Range Setup
{ 5, Channel Select, Data Select, (FFT Samples,) Data Begin, Data End, Step, K }
20020701
*1
Polynomial: Input constants in sequence, from
n
= 0 to 9.
*2
Mixed polynomial: Input constants in sequence from
m
= 4 to 1, and
n
= 0 to 5.
*3
Input of zero for constants can be skipped if all remaining constants are not used.
*4
Input 0 for constants that are not used.
•
When the conversion result of the “conversion equation” selected by Command 4 causes an overflow, the EA-200 sends a result of zero (0) to the calculator.