Technical specifications – Ion Science Hydrosteel 7000 User Manual
Page 40
Hydrosteel 7000 MANUAL
Ion Science Ltd
Page 40 of 43
Unrivalled Detection. www.ionscience.com
Technical specifications
Environmental
Operating Temperature
-10 to 50
C (-14 to 122
F)
Storage Temperature
-25 to 50
C (-13 to 140
F)
Zone 1/2 Safe Operating Temperature
-40 to 60
C (-40 to 140
F)
IP rating monitor
IP66
(note unit should not be mounted upside down)
IP66
(note unit should not be mounted upside down)
IP rating probe IP40, The weather resistance of the probe and hence the Hydrosteel system is dependant
on the probe installation. Caution; the probe must be sheltered from water and dust ingress as this will block
sample flow and prevent operation.
Mechanical
Monitor Dimensions & Mounting points monitor
Connector gland sizes for cable OD’s
1:
6-12 mm
2:
6-12 mm
4-20mA #1:
6-12 mm
4-20mA #2:
6-12 mm
Monitor Weight
~2.6 kg
5.75 lb
Monitor Materials
Glass filled polyester case
Electrical Power requirements:
PS1:
15-24 V @ 34 mA
PS2:
Spare. Contact Ion Science Ltd if power available for PS1 is not
sufficient due to local intrinsic safety requirement
4-20 mA #1:
9-30 V @ 2-20 mA (loop powered)
4-20 mA #2: