Hydrosteel 7000 MANUAL
Ion Science Ltd
Page 17 of 43
Unrivalled Detection. www.ionscience.com
Installation
Also check that the bands are fully parallel. NB: The probe can slide along the banding by loosening the
thumbscrews on the probe brackets, provided the banding is not too tight onto the pipe. Check the banding
buckles are a reasonable distance from the plate so that the banding runs flat against the steel surface and
then onto the collector plate. If the band buckles are very close to the collector edge the banding may not
pull the collector plate flat on to the surface with no or only a very small gap.
Step 4: Lock off the banding as normal, following the banding manufactures instructions. Wait 3-5 minutes!
Due to expansion of the banding attached to pipe at elevated temperatures, it may require tightening after
some time. Re-check probe tightness and positioning as per step 3 and retighten or refit banding as
necessary.
Step 5: Position the background probe so that it is sampling from the edge of the plate.