Hand over – Ion Science Hydrosteel 7000 User Manual
Page 31
Hydrosteel 7000 MANUAL
Ion Science Ltd
Page 31 of 43
Unrivalled Detection. www.ionscience.com
Commissioning
Hand over
1. With the permission of the owner, photograph the unit as installed. The photographs will provide a
documentary record of sound installation. Photographs should show the following.
Correct probe installation
Probe is securely mounted and safe from foreseeable damage
Sample conduit has been mounted securely and is safe from foreseeable damage
HY7K is mounted securely and is safe from foreseeable damage
All connections have been made securely to the HY7K and are safe from foreseeable damage
2. Produce hand-over document with the following details:
Customer details
Customers PO no
Hand-over date
Site or location reference
Equipment tag number or other reference
Add the photographs
Operation test results
First flux and pipe temperature measurements
3. Have the installation inspected and the hand-over the signed document.