ND 2100G GAGE-CHEK

Installation Instructions

en

17

4

Installation

NOTICE

Disturbances caused by lack of grounding or improper

grounding!

Never operate the product without proper ground
connection.
Connect the ground connector on the rear panel to
the central grounding point of the product.
Minimum cross section of the conductor: 6 mm

2

.

NOTICE

Risk of damage to internal components!

Do not engage or disengage any connecting elements
while the product is under power.

Depending on the product's equipment, the

installation may differ from the procedure

described in this chapter. If the Addendum

included with the product contains information

on the installation, the information given in

the Addendum takes precedence over the

information in this chapter.

The safety of any system incorporating the

use of this product is the responsibility of the

assembler or installer of the system.

The illustrations of pin layouts always show the

pin assignments of the connectors, and not of

the connections on the product.

Personnel requirement

The following steps are only to be performed by

qualified personnel!
For more information, see "Personnel

qualification", page 14.

4.1

Product overview

Rear panel

See A on the front fold-out page.

1

Power switch

2

Relay outputs

3

I/O interface

4

RS-232-C/V.24 port

5

Inputs for measuring devices

6

Power connector

7

Line fuse

The number and type of measuring device

connectors may vary from unit to unit.

Left side panel

The left side panel (as seen from the front) provides the

following connectors:
See B on the front fold-out page.

1

Speaker/headphone jack

2

USB Type A connector

3

RJ-45 jack for foot switch or remote keypad

4.2

Connecting power

WARNING

Risk of electric shock!
Improper grounding of electrical devices involves a risk of

serious injury or death by electric shock.

Always use 3-wire power cables.
Make sure the ground wire is correctly connected to
the ground of the building's electrical installations.

WARNING

Fire hazard from the use of power cables that do not

meet the minimum requirements!

Use only power cables that fulfill or exceed the
specified minimum requirements.

NOTICE

Surge suppressors limit the amplitude of potentially

damaging power line transients caused by electrical

machinery or lightning, and protect the product from

most power line transients that can corrupt system

memory or damage circuits.

Always connect the unit to power through a high-
quality surge suppressor.

Use the supplied power cable to connect the power
connector to a 3-wire grounded power outlet. For
information about the location of the power connector
on the rear panel, see "Rear panel", page 17.