Parameter

Settings/ Function

P40.2 COMP.

Select encoder compensation X2
No compensation

COMP. OFF

Nonlinear: up to 64
compensation points

NONL. COMP

Linear compensation

COMP. ON

P41.1 L.COMP.

Linear error compensation X1
– 99 999.9 < P41 < + 99 999.9 [µm/m]
Default setting: 0

Example: Find input value for P41
Displayed measuring length ............. L

a

= 620.000 mm

Actual length (measured, e.g.
with the VM 101 comparator
system from HEIDENHAIN) .............. L

t

= 619.877 mm

Length difference .............................

L = L

t

– L

a

= – 123 µm

Compensation factor k (= P41):
k =

L / L

a

= – 123 µm / 0.62 m ........ k = – 198.4 [µm/m]

P41.2 L.COMP.

Linear error compensation X2
– 99 999.9 < P41 < + 99 999.9 [µm/m]
Default setting: 0

Parameter

Settings/ Function

P31.1 S. PER.

Signal period of encoder X1
0.000 000 01 < P31 < 99 999.9999
Default setting: 10 µm

P31.2 S. PER.

Signal period of encoder X1
0.000 000 01 < P31 < 99 999.9999
Default setting: 10 µm

P33.1 COUNT

Counting mode X1
0-1-2-3-4-5-6-7-8-9

COUNT 0-1

0-2-4-6-8

COUNT 0-2

0-5

COUNT 0-5

P33.2 COUNT

Counting mode X2
0-1-2-3-4-5-6-7-8-9

COUNT 0-1

0-2-4-6-8

COUNT 0-2

0-5

COUNT 0-5

P38.1 DP POS

Decimal places X1

1)

1 / 2 / 3 / 4 / 5 / 6
(up to 8 for inch display)

P38.2 DP POS

Decimal places X2

1)

1 / 2 / 3 / 4 / 5 / 6
(up to 8 for inch display)

List of Operating Parameters

1)

Depends on signal period (P31) and measuring unit (P01)

P40.1 COMP.

Select encoder compensation X1
No compensation

COMP. OFF

Nonlinear: up to 64
compensation points

NONL. COMP

Linear compensation

COMP. ON