List of operating parameters – HEIDENHAIN ND 231 B v.1 User Manual
Page 19

19
Parameter
Settings/ Function
P40.2 COMP.
Select encoder compensation X2
No compensation
COMP. OFF
Nonlinear: up to 64
compensation points
NONL. COMP
Linear compensation
COMP. ON
P41.1 L.COMP.
Linear error compensation X1
– 99 999.9 < P41 < + 99 999.9 [µm/m]
Default setting: 0
Example: Find input value for P41
Displayed measuring length ............. L
a
= 620.000 mm
Actual length (measured, e.g.
with the VM 101 comparator
system from HEIDENHAIN) .............. L
t
= 619.877 mm
Length difference .............................
∆
L = L
t
– L
a
= – 123 µm
Compensation factor k (= P41):
k =
∆
L / L
a
= – 123 µm / 0.62 m ........ k = – 198.4 [µm/m]
P41.2 L.COMP.
Linear error compensation X2
– 99 999.9 < P41 < + 99 999.9 [µm/m]
Default setting: 0
Parameter
Settings/ Function
P31.1 S. PER.
Signal period of encoder X1
0.000 000 01 < P31 < 99 999.9999
Default setting: 10 µm
P31.2 S. PER.
Signal period of encoder X1
0.000 000 01 < P31 < 99 999.9999
Default setting: 10 µm
P33.1 COUNT
Counting mode X1
0-1-2-3-4-5-6-7-8-9
COUNT 0-1
0-2-4-6-8
COUNT 0-2
0-5
COUNT 0-5
P33.2 COUNT
Counting mode X2
0-1-2-3-4-5-6-7-8-9
COUNT 0-1
0-2-4-6-8
COUNT 0-2
0-5
COUNT 0-5
P38.1 DP POS
Decimal places X1
1)
1 / 2 / 3 / 4 / 5 / 6
(up to 8 for inch display)
P38.2 DP POS
Decimal places X2
1)
1 / 2 / 3 / 4 / 5 / 6
(up to 8 for inch display)
List of Operating Parameters
1)
Depends on signal period (P31) and measuring unit (P01)
P40.1 COMP.
Select encoder compensation X1
No compensation
COMP. OFF
Nonlinear: up to 64
compensation points
NONL. COMP
Linear compensation
COMP. ON
1)