List of operating parameters – HEIDENHAIN ND 231 B v.1 User Manual
Page 18

18
Parameter
Settings / Function
P00 CODE
Enter the code number 9 51 48 to change
the protected operating parameters
P01
Units of measurement
1)
Display in millimeters
MM
Display in inches
INCH
P06 DISPL.
Displayed information
Signals from encoder at X1
DISPLAY X1
Signals from encoder at X2
DISPLAY X2
Sum display X1 + X2
X1 ADD. X2
Difference display X2 – X2
X1 SUB. X2
P11 SCL
Scaling factor
Scaling factor on
SCALING ON
Scaling factor off
SCALNG. OFF
P12.1 SCL
Scaling factor X1
Enter a numerical value
0.100000 < P12 < 9.999999
Default setting: 1.000000
P12.2 SCL
Scaling factor X2
Enter a numerical value
0.100000 < P12 < 9.999999
Default setting: 1.000000
List of Operating Parameters
List of Operating Parameters
Parameter
Settings / Function
P17 SORT
Sorting and tolerance checking
Sorting and tol. checking ON
SORT ON
Sorting and tol. checking OFF
SORT OFF
P18 L.LIMIT
Lower limit for sorting
P19 U.LIMIT
Upper limit for sorting
P23 DISPL.
Display stop for measured value output
Concurrent display, no display freeze;
the display value is the current actual
value
DISPL. ACTL.
Frozen display; hold display until next
measured value output
DISPL. HOLD
Frozen/concurrent display; freeze display
as long as Pulse/Contact for measured
value output is present
DISPL. STOP
P30.1 DIR
Counting direction X1
Positive counting direction with
positive direction of traverse
DIRECT. POS
Negative counting direction with
positive direction of traverse
DIRECT. NEG
P30.2 DIR
Counting direction X2
Positive counting direction with
positive direction of traverse
DIRECT. POS
Negative counting direction with
positive direction of traverse
DIRECT. NEG