List of operating parameters – HEIDENHAIN ND 231 B v.1 User Manual

Page 18

background image

18

Parameter

Settings / Function

P00 CODE

Enter the code number 9 51 48 to change
the protected operating parameters

P01

Units of measurement

1)

Display in millimeters

MM

Display in inches

INCH

P06 DISPL.

Displayed information
Signals from encoder at X1

DISPLAY X1

Signals from encoder at X2

DISPLAY X2

Sum display X1 + X2

X1 ADD. X2

Difference display X2 – X2

X1 SUB. X2

P11 SCL

Scaling factor
Scaling factor on

SCALING ON

Scaling factor off

SCALNG. OFF

P12.1 SCL

Scaling factor X1
Enter a numerical value
0.100000 < P12 < 9.999999
Default setting: 1.000000

P12.2 SCL

Scaling factor X2
Enter a numerical value
0.100000 < P12 < 9.999999
Default setting: 1.000000

List of Operating Parameters

List of Operating Parameters

Parameter

Settings / Function

P17 SORT

Sorting and tolerance checking
Sorting and tol. checking ON

SORT ON

Sorting and tol. checking OFF

SORT OFF

P18 L.LIMIT

Lower limit for sorting

P19 U.LIMIT

Upper limit for sorting

P23 DISPL.

Display stop for measured value output
Concurrent display
, no display freeze;
the display value is the current actual
value

DISPL. ACTL.

Frozen display; hold display until next
measured value output

DISPL. HOLD

Frozen/concurrent display; freeze display
as long as Pulse/Contact for measured
value output is present

DISPL. STOP

P30.1 DIR

Counting direction X1
Positive counting direction with
positive direction of traverse

DIRECT. POS

Negative counting direction with
positive direction of traverse

DIRECT. NEG

P30.2 DIR

Counting direction X2
Positive counting direction with
positive direction of traverse

DIRECT. POS

Negative counting direction with
positive direction of traverse

DIRECT. NEG