Watson-Marlow 505Du User Manual
Page 21

21
T H E N
P R E S S
E N T E R
then press Enter
A P P L Y
+
8 . 0 0 M A
Apply 8mA across pin 14+ and pin 17-
T H E N
P R E S S
E N T E R
then press Enter
T R I M
C O M P L E T E
Confirmation message
B A U D - 9 6 0 0
T R I M
>
The SETUP menu returns
Signal calibration is now complete.
The pump can be set so that after the mains supply is interrupted the pump
automatically restarts in the same state as before the interruption or, that after
the power is reconnected the pump will remain stopped.
If you press Enter with the cursor at OFF, following a power interruption the
pump will return to the main menu, and await instructions. If you Step to ON, then
press Enter, the pump will automatically re-start following a power interruption.
The factory default is Auto-restart OFF - the pump returns to the main menu.
To change the Auto-restart state:
A U T O S T A R T - O F F
>
Current state displayed. Press Enter
A U T O - S T A R T
W H E N
P O W E R E D
U P ?
Explanatory message
A U T O S T A R T
O N
O F F
Step, and press Enter
A U T O S T A R T - O N
>
Confirmation. Step or press Main menu
The state of the pump may be monitored by a system of two 5V Hi Lo signals by
connecting to the 25D remote socket on the pump rear panel. This feature can
Autostart
Strobe