Watson-Marlow 505Du User Manual
Page 18

18
4
MAX
25
17
Calibration
If it is required to display the flow rate in ml/min the processor must be informed
which pumphead and tubing is being used. If your application requires accurate
indication of flow rate you should then proceed to a calibration dose.
In the main menu, Step to:
C A L I B
S E T U P
<
Press Enter at CALIB
C U R R E N T
R E - C A L I B
Press Enter to view current calibration
The display will show for a three seconds the pumphead, ml per revolution, tube
size and the speed at which the pump is set to run. The screen will then display:
C U R R E N T
R E - C A L I B
Step, Enter to recalibrate
If you have selected to enter a new head or tube, the display will now show:
H E A D
C A L - D O S E
Press Enter at HEAD
The first three pumpheads from the range are shown:
5 0 1 R L
5 0 5 L
3 0 3
>
Step to the pumphead fitted and press Enter
All the tube sizes which can be used in the pumphead selected are shown.
0 . 5
0 . 8
1 . 6
3 . 2
>
Step to the tubing fitted and press Enter
The display will now confirm your selection of pumphead and tubing:
5 0 1 R L
3 . 2 M M
Confirmation screen
Then the display will move on to request if a calibration dose is required.