7.7.5 If the displayed value is less than

“200” (decimal point

not shown), a more accurate reading may be obtained
by setting the RANGE switch to a lower range. Rotate
the RANGE switch to the next lower position and
observe the reading on the LCD display. If the RANGE
switch is already on the lowest position (i.e. 200
Ohms), no greater measurement resolution can

be obtained.


Notes:

a) The 2M, 20M, and 200M ranges require several seconds to stabilize.

b) The 200M range reads about 10 LSD high. To obtain an accurate

reading, short the test leads together and observe the residual

reading (usually 8 to 12 LSD). For the best accuracy, subtract

observed reading from subsequent measurements.

c) To obtain the most accurate reading on the 200

Ω range, short the

test leads together and note the

“residual resistance” reading.

It is typically less than 0.5

Ω . Subtract this value from subsequent

readings on the 200

Ω range.