beautypg.com

Tecfluid FLOMID-MX User Manual

Page 18

background image

18

7 STATUS

INDICATION


During the normal working mode, some symbols can be displayed on the screen which indicate

the status of the instrument.

If the flow rate has risen above the maximum fluid speed (11 m/s) an asterisk (*) will appear in

front of the flow rate value.

If the total volume counter has over flown (>9.999.999 or < 0 units) an asterisk (*) will appear in

front total volume value.


8

EXAMPLES OF USEFUL CALCULATIONS


Measurement error corrections

The calibration of flow meters is accomplished using, for the liquid, water at 20

o

C thus obtaining

a calibration for a liquid of density 1 kg/l and viscosity of 1 mPas. If the flow meter is used with a liquid of
other characteristics from the above specified o for reasons of turbulences in the flow, measurement
errors can be induced.

To correct these types of errors we can modify the value of the body calibration factor Fc

introduced in the configuration of the instrument.

Example 1 - Shortage of volume

If we have a flow meter body which specifies Fc = 0.985, and when we check the volume of a

batch, we find that instead of having 100 litres as programmed, we only have 95 litres (5% less) and we
must apply the following correction:

Fcn = Calibration Factor (new)

= ? (1.03684)

Fc = Calibration Factor (original)

= 0.985

V = Expected Volume

= 100

Vr = Real Volume

= 95


Example 2 - Excess of volume


If we have a flow meter body which specifies Fc = 0.985, and when we check the volume of a

batch, we find that instead of having 100 litres as programmed, we only have 105 litres (5% more) and
we must apply the following correction:

Fcn = Calibration Factor (new)

= ? (0.89545)

Fc = Calibration Factor (original)

= 0.985

V = Expected Volume

= 100

Vr = Real Volume

= 105

Change of units of measurement

In some cases we need to change the measurement units for batching, for example, instead of

working in litres we need to work in kilograms. In this case we will need to know the density of the liquid
(ρ).

To change from litres to kilos we must divide the body calibration factor Fc by the density of the

liquid to obtain the new factor for programming the instrument. For example, if the liquid has a density of
0.9 and the body calibration factor Fc = 0.985, and we must batch in kilos; we will have to introduce Fc =
1.09444 in the basic configuration to be able to preset directly in kilos.

Fcd

= Calibration Factor (new density)

Fc

= Calibration Factor (original)

ρ

= Density of the liquid in kg/litre

V

V

F

=

F

r

c

cn

V

V

F

=

F

r

c

cn

ρ

F

=

F

c

cd