Tecfluid FLOMID-MX User Manual
Page 10

10
Once the minimum flow has been configured the set-up returns to the main menu showing
flow rate set-up.
Pressing
the
push button the next item on the main menu is displayed
4.2 Pulse
output
P
u l s e
o u t p u t
To enter this submenu the
push button is pressed.
Two possibilities exist for this set-up and are selected using the
push button.
P
u l s e
o u t p u t
F r e q u e n c y
H z
P
u l s e
o u t p u t
P u l s e s / u n i t
Pressing the
push button the option displayed is selected.
4.2.1 Frequency
(Hz)
The instrument will give a frequency output proportional to the flow rate for flow speeds between
0 and 10 m/s, and it will be 1000 Hz when the flow rate coincides with Qnom. (see 4.1.4). The frequency
will be zero when the flow rate is below the minimum flow rate (see 4.1.6). The minimum frequency
output in this mode is 10 Hz.
4.2.2
Pulses per measuring unit
P u l s e s / u n I t
>
0
0 . 5 0 0
The pulse output will give the pulses corresponding to the volume flowing through the sensor.
This mode is useful for PLC and counter inputs.
NOTE: In the pulse/unit mode the maximum frequency is 2 Hz. If the flow rate is higher than that
required to give 2 Hz, then the output frequency will drop to zero.
NOTE: In the event of using electromechanical counters make sure that the counter will accept the
maximum frequency that can occur at the output.
Once the pulse output has been configured, we can choose the pulse width between two
options: A 50% duty cycle square wave or a pulse width of 65 ms with the output active.
P
u l s e w i d t h
5 0 %