beautypg.com

Tecfluid FLOMID-MX User Manual

Page 17

background image

17


By pressing the key

, the state of the digital outputs will change. By pressing the key

,

the state of the frequency output will change. By pressing the key

, the analog output will change,

from 4 to 20 mA or vice versa. The state of the outputs is shown on the screen.

0 1 4

O

u

t P u l

m A


With the key

, we will leave the test.

4.8 Version

In the last configuration screen, the version of the software is shown.

5 BATCHING


5.1 Preset

entry

From the “standby” screen the

push button should be pressed to enter the batching quantity

entry.

P r e s e t

> 0 0 0 1 0 0 0 m 3


A value of 0000000 will not be admitted


5.2 Batching

To start the batching process, at least one of the two outputs must be configured for batching

(see part 4.4.1 and 4.4.2) whilst in the normal working mode, push the

push button. In this moment

the partial volume counter will be reset and corresponding output (as configured in 4.4) will be activated.
The screen shows the word "Batching" and the partial counter. The batching process will begin and the
instrument will count the volume that passes through the sensor and add it to the partial volume counter.
When the partial volume counter reaches the preset value the output will deactivate and the batching
process finishes. The screen will show the word "Partial" instead of "Batching".

The total volume counter will act as configured in 4.6.2.

If during the batching cycle the

push button is pressed, the output will be deactivated, the

batching process will be stopped, although the batch is not finished. In this case the screen will show the
word "Partial" instead of "Batching". Assuming that the flow stops when not batching, we can see at what
stage the batch was stopped (partial volume counter). If we need to finish a batch which has been
stopped then we must preset the difference between the original quantity and the partial counter when
the batch was stopped. If there is a power fail during a batching process, the effect will be the same as
having pressed the

push button.


For a remote control, to start the batching process there is a reset input at the rear terminal

block (see 2.4). This input works exactly the same as the

push button.

6

TOTAL VOLUME COUNTER


The total volume counter indicates the total volume that has passed through the sensor as

configured in part 4.6.2. This counter is independent of the partial volume counter used in the batching
process.

To reset the total volume counter to zero at first we must be in the standby mode and then press

the

and

push buttons at the same time.

If the total volume counter is increasing and reaches a value of 9.999.999 units it will overflow

on the next count and go to zero. If it is configured to count down and reaches zero on the next count it
will go to 9.999.999.